  • RE: EasyStat Na Hct

    Dear Tobias,

     

    Thank you very much for the print outs.

     

    Based on the print outs the analyzer is working well.

    The percentage of red blood cells to the total blood volume is defined as hematocrit. It is quantified by measuring the electrical impedance of blood sample. Two standard solutions are used to calibrate the hematocrit sensor. The analyzer then measures the electrical impedance of the blood sample to obtain the hematocrit value. Next, the hematocrit value obtained is corrected for the concentration of the sodium ion.( Please see the attached file from user manual)

    So you always need to have a working sodium electrode ( with the good calibration results and QC results ) to have the good results for HCT.

     

    Sorry for the confusion Tobias.

     

    Best Regards,

     

    Hanna Parhizkar | Technical Support  | hparhizkar@medicacorp.com

    Medica Corporation | 5 Oak Park Drive, Bedford, MA 01730 | www.medicacorp.com

     

    +1 781 275 4892 Main

    +1 781 541 7403 Direct                    

    +1 781 275 2731 Fax

     

     

     

     

     

     

     

     

     

     

    From: Tobias Pfeifer <tobias.pfeifer@fameco.eu>
    Sent: Tuesday, October 22, 2019 4:14 AM
    To: MEDICA TechnicalSupport <medicatechnicalsupport@fameco.odoo.com>
    Subject: EasyStat Na Hct

     

    Dear Hanna,
    Here are the printouts I got from the Na Hct problem. 
    Best regards 
    Tobias


    --

    Image removed by sender.

    Tobias Pfeifer | Geschäftsführung

    T: +49 89 248805810  | F: +49 89 248805819

    E: tobias.pfeifer@fameco.eu | W: https://fameco.eu/

    FAMECO GmbH | Geschäftsführer: Tobias Pfeifer, Thomas Andersen

    Neumarkter Str. 34, 81673 München

    USt-IdNr. DE323800861 | Amtsgericht München HRB 248816

    The company accepts no liability for the content of this email, or for the consequences of any actions taken on the basis of the information provided, unless that information is subsequently confirmed in writing. If you are not the intended recipient you are notified that disclosing, copying, distributing or taking any action in reliance on the contents of this information is strictly prohibited.

     

    Sent by FAMECO GmbH using Odoo.


    durch Hanna Parhizkar - 01:56 - 22 Okt. 2019
  • EasyStat Na Hct

    Dear Hanna,
    Here are the printouts I got from the Na Hct problem. 
    Best regards 
    Tobias


    durch Tobias Pfeifer - 03:13 - 22 Okt. 2019
  • RE: ES Na & Hct

    Dear Tobias,

     

    Thank you for your email.

     

    1. Could you please let us know what is the serial number on this analyzer?

    2. Could you please send us the calibration print out?

    3. You mentioned that whenever the Na is not working the Hct is not working either. Could you please send us a print out?

    4. Which kind of sample do they test?

    5. Which kind of sample tubing do they use?

    6. What is the part number for the daily cleaner? For how long do they use this fresh daily cleaner after making a new one?

    7. No, Na and Hct don’t have any connection.

     

    Looking forward to hearing back from you!

     

    Best regards,

     

    Hanna Parhizkar | Technical Support  | hparhizkar@medicacorp.com

    Medica Corporation | 5 Oak Park Drive, Bedford, MA 01730 | www.medicacorp.com

     

    +1 781 275 4892 Main

    +1 781 541 7403 Direct                    

    +1 781 275 2731 Fax

     

     

    Signature031319_96dpi400px-01

     

     

     

    From: Tobias Pfeifer <tobias.pfeifer@fameco.eu>
    Sent: Wednesday, October 16, 2019 7:53 AM
    To: MEDICA TechnicalSupport <medicatechnicalsupport@fameco.odoo.com>
    Subject: ES Na & Hct

     

    Dear Hanna,

    I have a doctor who claims that whenever the Na is not working also the Hct is not working.
    He needs Hct but doesnt need Na at all.
    Is there a connection?
    best regards
    Tobias


    --

    Image removed by sender.

    Tobias Pfeifer | Geschäftsführung

    T: +49 89 248805810  | F: +49 89 248805819

    E: tobias.pfeifer@fameco.eu | W: https://fameco.eu/

    FAMECO GmbH | Geschäftsführer: Tobias Pfeifer, Thomas Andersen

    Neumarkter Str. 34, 81673 München

    USt-IdNr. DE323800861 | Amtsgericht München HRB 248816

    The company accepts no liability for the content of this email, or for the consequences of any actions taken on the basis of the information provided, unless that information is subsequently confirmed in writing. If you are not the intended recipient you are notified that disclosing, copying, distributing or taking any action in reliance on the contents of this information is strictly prohibited.

     

    Sent by FAMECO GmbH using Odoo.


    durch Hanna Parhizkar - 03:16 - 16 Okt. 2019
  • RE: ES & EBG packs policy

    Dear Tobias,

     

    The shelf life for the ES and EBG reagent modules are one year from the date of manufacture. Please make sure to use them before this time.

    Es and EBG reagent modules have dallas chip. When you install these reagent modules, the dallas chip reads the date and analyzer serial number ect.

    When you install the EBG reagent module you can use it for 45 days. You can use ES reagent module for 35 days when you install it.

     

    Regarding your questions:

     

    1. No, you can only use it until the expiration date. When it is expired you can’t use it any more. If you install one reagent module on it’s last day you can’t use it the day after or for 35 or 45 days because it will give you “expired reagent module” message.

    2. Unfortunately we don’t have this option on our reagent modules and you can swap them between analyzers.

    But you can see from the print out that this reagent module was installed on which EBG analyzer for the first time:

    1. Please press 5 for second menu

    2. Press 4 for stored information

    3. Press 5 for reagent module

    4. Press 1 for status. Here you can see the analyzer’s serial number that the reagent module was installed for the first time.

     

    Hope it helps.

     

    Please contact us if you have any questions.

     

    Best regards,

     

    Hanna Parhizkar | Technical Support  | hparhizkar@medicacorp.com

    Medica Corporation | 5 Oak Park Drive, Bedford, MA 01730 | www.medicacorp.com

     

    +1 781 275 4892 Main

    +1 781 541 7403 Direct                    

    +1 781 275 2731 Fax

     

     

    Signature031319_96dpi400px-01

     

     

     

     

     

     

     

    From: Tobias Pfeifer <tobias.pfeifer@fameco.eu>
    Sent: Wednesday, October 16, 2019 7:52 AM
    To: MEDICA TechnicalSupport <medicatechnicalsupport@fameco.odoo.com>
    Subject: ES & EBG packs policy

     

    Dear Hanna,

    two questions regarding the packs:

    1. will the pack if installed on the last day of it's expiry (IN date) work through the whole 45 days, or will it be cut of to the date thats on the box?
    2. where can I find guidance on how the packs are protected against swapping from one EBG to another EBG for example?

    best regards
    Tobias


    --

    Image removed by sender.

    Tobias Pfeifer | Geschäftsführung

    T: +49 89 248805810  | F: +49 89 248805819

    E: tobias.pfeifer@fameco.eu | W: https://fameco.eu/

    FAMECO GmbH | Geschäftsführer: Tobias Pfeifer, Thomas Andersen

    Neumarkter Str. 34, 81673 München

    USt-IdNr. DE323800861 | Amtsgericht München HRB 248816

    The company accepts no liability for the content of this email, or for the consequences of any actions taken on the basis of the information provided, unless that information is subsequently confirmed in writing. If you are not the intended recipient you are notified that disclosing, copying, distributing or taking any action in reliance on the contents of this information is strictly prohibited.

     

    Sent by FAMECO GmbH using Odoo.


    durch Hanna Parhizkar - 03:00 - 16 Okt. 2019
  • ES Na & Hct

    Dear Hanna,

    I have a doctor who claims that whenever the Na is not working also the Hct is not working.
    He needs Hct but doesnt need Na at all.
    Is there a connection?
    best regards
    Tobias


    durch Tobias Pfeifer - 06:52 - 16 Okt. 2019
  • ES & EBG packs policy

    Dear Hanna,

    two questions regarding the packs:

    1. will the pack if installed on the last day of it's expiry (IN date) work through the whole 45 days, or will it be cut of to the date thats on the box?
    2. where can I find guidance on how the packs are protected against swapping from one EBG to another EBG for example?

    best regards
    Tobias


    durch Tobias Pfeifer - 06:51 - 16 Okt. 2019
  • RE: EBG BD

    Dear Tobias,

     

    Thank you for your email.

     

    The top bubble detector and the bottom bubble detector are different. ( Please see the attached pictures for your information)

    Depending on the error that you get you might need one of them or both of them. I would recommend to have both of them with you when you go for repair.

    Please see the attached tech tip for your information.

     

    Please contact us if you have any questions.

     

    Best regards,

     

    Hanna Parhizkar | Technical Support  | hparhizkar@medicacorp.com

    Medica Corporation | 5 Oak Park Drive, Bedford, MA 01730 | www.medicacorp.com

     

    +1 781 275 4892 Main

    +1 781 541 7403 Direct                    

    +1 781 275 2731 Fax

     

     

    Signature031319_96dpi400px-01

     

     

     

    From: Tobias Pfeifer <tobias.pfeifer@fameco.eu>
    Sent: Tuesday, October 15, 2019 3:25 PM
    To: MEDICA TechnicalSupport <medicatechnicalsupport@fameco.odoo.com>
    Subject: EBG BD

     

    Dear Hanna,
    What's the difference between 6801 and 6802 BD Kit Top and bottom, do I need both for 1 repair?
    Best regards 
    Tobias


    --

    Image removed by sender.

    Tobias Pfeifer | Geschäftsführung

    T: +49 89 248805810  | F: +49 89 248805819

    E: tobias.pfeifer@fameco.eu | W: https://fameco.eu/

    FAMECO GmbH | Geschäftsführer: Tobias Pfeifer, Thomas Andersen

    Neumarkter Str. 34, 81673 München

    USt-IdNr. DE323800861 | Amtsgericht München HRB 248816

    The company accepts no liability for the content of this email, or for the consequences of any actions taken on the basis of the information provided, unless that information is subsequently confirmed in writing. If you are not the intended recipient you are notified that disclosing, copying, distributing or taking any action in reliance on the contents of this information is strictly prohibited.

     

    Sent by FAMECO GmbH using Odoo.





    Sent from my iPhone







    Sent from my iPhone

    durch Hanna Parhizkar - 03:51 - 15 Okt. 2019
  • EBG BD

    Dear Hanna,
    What's the difference between 6801 and 6802 BD Kit Top and bottom, do I need both for 1 repair?
    Best regards 
    Tobias


    durch Tobias Pfeifer - 02:24 - 15 Okt. 2019
  • RE: Cost per Test

    Dear Tobias,

     

     

    I forwarded your email to the appropriate person. He is working on these questions. As soon as he sends me the answers, I will forward them to you.

     

    Thank you for your patience.

     

    Best regards,

     

    Hanna Parhizkar | Technical Support  | hparhizkar@medicacorp.com

    Medica Corporation | 5 Oak Park Drive, Bedford, MA 01730 | www.medicacorp.com

     

    +1 781 275 4892 Main

    +1 781 541 7403 Direct                    

    +1 781 275 2731 Fax

     

     

    Signature031319_96dpi400px-01

     

     

     

     

    From: Tobias Pfeifer <tobias.pfeifer@fameco.eu>
    Sent: Tuesday, October 8, 2019 2:30 PM
    To: MEDICA TechnicalSupport <medicatechnicalsupport@fameco.odoo.com>
    Subject: Cost per Test

     

    Dear Hanna,
    any news on the QC usage?
    Who could I contact for some general guidance on the supplied XLS calculating worksheets?
    best regards
    Tobias


    --

    Image removed by sender.

    Tobias Pfeifer | Geschäftsführung

    T: +49 89 248805810  | F: +49 89 248805819

    E: tobias.pfeifer@fameco.eu | W: https://fameco.eu/

    FAMECO GmbH | Geschäftsführer: Tobias Pfeifer, Thomas Andersen

    Neumarkter Str. 34, 81673 München

    USt-IdNr. DE323800861 | Amtsgericht München HRB 248816

    The company accepts no liability for the content of this email, or for the consequences of any actions taken on the basis of the information provided, unless that information is subsequently confirmed in writing. If you are not the intended recipient you are notified that disclosing, copying, distributing or taking any action in reliance on the contents of this information is strictly prohibited.

     

    Sent by FAMECO GmbH using Odoo.


    durch Hanna Parhizkar - 09:41 - 9 Okt. 2019
  • Cost per Test

    Dear Hanna,
    any news on the QC usage?
    Who could I contact for some general guidance on the supplied XLS calculating worksheets?
    best regards
    Tobias


    durch Tobias Pfeifer - 01:29 - 8 Okt. 2019
  • RE: EasyLyte QC usage

    Dear Tobias,

     

    I will forward your email to the appropriate person. As soon as I get the answer I will forward it to you.

     

    Best regards,

     

    Hanna Parhizkar | Technical Support  | hparhizkar@medicacorp.com

    Medica Corporation | 5 Oak Park Drive, Bedford, MA 01730 | www.medicacorp.com

     

    +1 781 275 4892 Main

    +1 781 541 7403 Direct                    

    +1 781 275 2731 Fax

     

     

    Signature031319_96dpi400px-01

     

     

     

    From: Tobias Pfeifer <tobias.pfeifer@fameco.eu>
    Sent: Monday, October 7, 2019 9:36 AM
    To: MEDICA TechnicalSupport <medicatechnicalsupport@fameco.odoo.com>
    Subject: EasyLyte QC usage

     

    Hey, 
    we have a new potential customer for a EL Expand; 
    they are doing 600 samples a week;
    they want to do QC everyday - is there a tool to calculate amounts and costs for the QCs?



     
    Sodium/Potassium/Chloride/Lithium Analyzer



    800 mL Reagent Pack Life Cost Per Test

    Samples Total(mL) Pack Total Samples Reagent Total Cost/ Total Cost/
    per 5day Std A Life Samples per 5day Cost per Sample * Test **
    Week Usage/Week (weeks) per Pack Week Sample

    135   98,8   8,1   1093     135   ,16 ,39   $0,10  
    100   80,6   9,9   993     100   ,17 ,49   $0,12  
    50   54,6   14,7   733     50   ,24 ,86   $0,22  
    40   49,4   16,2   648     40   ,27 1,05   $0,26  
    35   46,8   17,1   598     35   ,29 1,18   $0,30  
    30   44,2   18,1   543     30   ,32 1,36   $0,34  
    25   41,6   19,2   481     25   ,36 1,61   $0,40  
    20   39,0   20,5   410     20   ,42 1,98   $0,50  
    15   36,4   22,0   330     15   ,52 2,61   $0,65  

    Enter Samples per 5day Week* Enter Samples per 5day Week
    150   106,6   7,5   1126     150   ,15 ,39   ,10
    *Use for samples greater than 135 * Includes all reagents, maintenance, and supply
      items for 1 year
    ** 1 Sample = 3 Tests, Na,K and Li

    Assumptions Yearly Supply Requirements
    Samples per 5day Week, less than 135
    Reagent Pack Volume (mL) 800 Quantity** Description Price Total
    EasyLyte Price 3970,$00 2,0 Maintenance Kits 75,$00 150,$00
    Reimbursement/Sample 12,$81 0,7 Na Electrode 213,00 149,10
    Reagent Pack Cost 173,$00 1,3 K Electrode 190,00 247,00
    1,3 Li Electrode 250,00 325,00
    Standard A Usage (mL) 1,3 Cl Electrode 190,00 247,00
    Sample Analysis ,52 4,0 Disp. Ref. Electrode 95,00 380,00
    1st Calibration/day 1,58 2,0 Daily Cleaning Solution 40,00 80,00
    2nd, 3rd, 4th Cals ,50 2,0 Packages Printer Paper 23,00 46,00
    Sip ,04 1624,$10
    Cleaner ,32 ** Electrode usage based on typical useful electrode life
    Yearly Supply Requirements
    Samples per 5day Week, less than 135
    Base Standard A Usage Quantity*** Description Price Total
    7 Calibrations/week 13,56 2,0 Maintenance Kits 75,$00 150,$00
    5 Cleanings/week 1,60 0,8 Na Electrode 213,00 170,40
    336 Sips per week 13,44 1,5 K Electrode 190,00 285,00
    Total 28,60 1,5 Li Electrode 250,00 375,00
    1,5 Cl Electrode 190,00 285,00
    4,0 Disp. Ref. Electrode 95,00 380,00
    3,0 Daily Cleaning Solution 40,00 120,00
    3,0 Packages Printer Paper 23,00 69,00
    1834,$40
    ***Calculations derived from manual entry of sample volume



    best regards
    Tobias


    --

    Image removed by sender.

    Tobias Pfeifer | Geschäftsführung

    T: +49 89 248805810  | F: +49 89 248805819

    E: tobias.pfeifer@fameco.eu | W: https://fameco.eu/

    FAMECO GmbH | Geschäftsführer: Tobias Pfeifer, Thomas Andersen

    Neumarkter Str. 34, 81673 München

    USt-IdNr. DE323800861 | Amtsgericht München HRB 248816

    The company accepts no liability for the content of this email, or for the consequences of any actions taken on the basis of the information provided, unless that information is subsequently confirmed in writing. If you are not the intended recipient you are notified that disclosing, copying, distributing or taking any action in reliance on the contents of this information is strictly prohibited.

     

    Sent by FAMECO GmbH using Odoo.


    durch Hanna Parhizkar - 01:27 - 7 Okt. 2019
  • EasyLyte QC usage

    Hey, 
    we have a new potential customer for a EL Expand; 
    they are doing 600 samples a week;
    they want to do QC everyday - is there a tool to calculate amounts and costs for the QCs?



     
    Sodium/Potassium/Chloride/Lithium Analyzer



    800 mL Reagent Pack Life Cost Per Test

    Samples Total(mL) Pack Total Samples Reagent Total Cost/ Total Cost/
    per 5day Std A Life Samples per 5day Cost per Sample * Test **
    Week Usage/Week (weeks) per Pack Week Sample

    135   98,8   8,1   1093     135   ,16 ,39   $0,10  
    100   80,6   9,9   993     100   ,17 ,49   $0,12  
    50   54,6   14,7   733     50   ,24 ,86   $0,22  
    40   49,4   16,2   648     40   ,27 1,05   $0,26  
    35   46,8   17,1   598     35   ,29 1,18   $0,30  
    30   44,2   18,1   543     30   ,32 1,36   $0,34  
    25   41,6   19,2   481     25   ,36 1,61   $0,40  
    20   39,0   20,5   410     20   ,42 1,98   $0,50  
    15   36,4   22,0   330     15   ,52 2,61   $0,65  

    Enter Samples per 5day Week* Enter Samples per 5day Week
    150   106,6   7,5   1126     150   ,15 ,39   ,10
    *Use for samples greater than 135 * Includes all reagents, maintenance, and supply
      items for 1 year
    ** 1 Sample = 3 Tests, Na,K and Li

    Assumptions Yearly Supply Requirements
    Samples per 5day Week, less than 135
    Reagent Pack Volume (mL) 800 Quantity** Description Price Total
    EasyLyte Price 3970,$00 2,0 Maintenance Kits 75,$00 150,$00
    Reimbursement/Sample 12,$81 0,7 Na Electrode 213,00 149,10
    Reagent Pack Cost 173,$00 1,3 K Electrode 190,00 247,00
    1,3 Li Electrode 250,00 325,00
    Standard A Usage (mL) 1,3 Cl Electrode 190,00 247,00
    Sample Analysis ,52 4,0 Disp. Ref. Electrode 95,00 380,00
    1st Calibration/day 1,58 2,0 Daily Cleaning Solution 40,00 80,00
    2nd, 3rd, 4th Cals ,50 2,0 Packages Printer Paper 23,00 46,00
    Sip ,04 1624,$10
    Cleaner ,32 ** Electrode usage based on typical useful electrode life
    Yearly Supply Requirements
    Samples per 5day Week, less than 135
    Base Standard A Usage Quantity*** Description Price Total
    7 Calibrations/week 13,56 2,0 Maintenance Kits 75,$00 150,$00
    5 Cleanings/week 1,60 0,8 Na Electrode 213,00 170,40
    336 Sips per week 13,44 1,5 K Electrode 190,00 285,00
    Total 28,60 1,5 Li Electrode 250,00 375,00
    1,5 Cl Electrode 190,00 285,00
    4,0 Disp. Ref. Electrode 95,00 380,00
    3,0 Daily Cleaning Solution 40,00 120,00
    3,0 Packages Printer Paper 23,00 69,00
    1834,$40
    ***Calculations derived from manual entry of sample volume



    best regards
    Tobias


    durch Tobias Pfeifer - 08:35 - 7 Okt. 2019
  • RE: EasyLyte -2- claims

    Dear Tobias,

     

    Thank you for the failure report.

     

    On the EL EXPAND analyzer, every time that you have shifts in QC or high slope failure ( Calibration value ), please change the reference electrode and send us the reference electrode lot number, calibration results, QC data. We will send you a replacement for the reference electrode and make an RGA for you.

    We would recommend to run the daily cleaner in the morning, do calibration and run the QC materials. If they are good and within the normal ranges, run the patient samples.

     

    I processed the replacements for you. If you have any questions or concerns regarding the shipment of this order, please contact our Customer Service directly at : CustomerService@medicacorp.com

    We would like to get the reference electrode back to Medica for further investigations. I made the RGA 192085 for you. Please use our FedEx account number 471164747 and forward it to the address below:

     

    Medica Corporation

    Attn: Hanna Parhizkar ( RGA 192085 )

    3 Oak Park Drive

    Bedford, MA 01730

    Tel: 781-275-4892

     

    Please make sure this RGA number is visible on the external shipping crate. Please send us the tracking number when it is ready.

     

    Best regards,

     

    Hanna Parhizkar | Technical Support  | hparhizkar@medicacorp.com

    Medica Corporation | 5 Oak Park Drive, Bedford, MA 01730 | www.medicacorp.com

     

    +1 781 275 4892 Main

    +1 781 541 7403 Direct                    

    +1 781 275 2731 Fax

     

     

    Signature031319_96dpi400px-01

     

     

     

    From: Tobias Pfeifer <tobias.pfeifer@fameco.eu>
    Sent: Wednesday, October 2, 2019 2:38 AM
    To: MEDICA TechnicalSupport <medicatechnicalsupport@fameco.odoo.com>
    Subject: EasyLyte -2- claims

     

    Dear Hanna,
    here are 2 claims for EasyLyte Electrodes after we finished troubleshooting and exchanging parts.
    best regards
    Tobias


    --

    Image removed by sender.

    Tobias Pfeifer | Geschäftsführung

    T: +49 89 248805810  | F: +49 89 248805819

    E: tobias.pfeifer@fameco.eu | W: https://fameco.eu/

    FAMECO GmbH | Geschäftsführer: Tobias Pfeifer, Thomas Andersen

    Neumarkter Str. 34, 81673 München

    USt-IdNr. DE323800861 | Amtsgericht München HRB 248816

    The company accepts no liability for the content of this email, or for the consequences of any actions taken on the basis of the information provided, unless that information is subsequently confirmed in writing. If you are not the intended recipient you are notified that disclosing, copying, distributing or taking any action in reliance on the contents of this information is strictly prohibited.

     

    Sent by FAMECO GmbH using Odoo.

    Dear Tobias,

     

    Your warranty order #TSW-192085 has been processed. Attached is a summary of your order. You will receive a formal order acknowledgement from our Customer Service Team shortly.

     

    Best Regards,

     

    Hanna Parhizkar

     

     


    durch Hanna Parhizkar - 01:16 - 2 Okt. 2019
  • EasyStat BD & Calibration

    Dear Hanna,
    please see this most recent printouts - how come BD Test is good - Calibration than bad, then good - it's working as of this moment but nobody trusts the instrument anymore until we can find some explanation and final solution.
    best regards
    Tobias


    durch Tobias Pfeifer - 04:56 - 2 Okt. 2019
  • EasyLyte -2- claims

    Dear Hanna,
    here are 2 claims for EasyLyte Electrodes after we finished troubleshooting and exchanging parts.
    best regards
    Tobias


    durch Tobias Pfeifer - 01:38 - 2 Okt. 2019
  • EasysStat SN 1117031005

    Dear Hanna,
    here are the printouts - after cleaning the BD and the valve module also with bleach
    best regards
    Tobias


    durch Tobias Pfeifer - 01:13 - 2 Okt. 2019
  • Re: EasyStat
    Dear Tobias,

    Unfortunately I didn’t get the print outs.
    Could you please re-send the print outs?

    Thank you very much!

    Best Regards,
    Hanna
    From: Tobias Pfeifer <tobias.pfeifer@fameco.eu>
    Sent: Tuesday, October 1, 2019 6:26:49 PM
    To: MEDICA TechnicalSupport
    Subject: EasyStat
     

    Dear Hanna,

    I had sent the printouts in a separate mail 5h ago, did you get it?

    The SN is 1117031005

    Best regards 
    Tobias


    --

    Tobias Pfeifer | Geschäftsführung

    T: +49 89 248805810  | F: +49 89 248805819

    E: tobias.pfeifer@fameco.eu | W: https://fameco.eu/

    FAMECO GmbH | Geschäftsführer: Tobias Pfeifer, Thomas Andersen

    Neumarkter Str. 34, 81673 München

    USt-IdNr. DE323800861 | Amtsgericht München HRB 248816

    The company accepts no liability for the content of this email, or for the consequences of any actions taken on the basis of the information provided, unless that information is subsequently confirmed in writing. If you are not the intended recipient you are notified that disclosing, copying, distributing or taking any action in reliance on the contents of this information is strictly prohibited.


    Sent by FAMECO GmbH using Odoo.


    durch Hanna Parhizkar - 07:58 - 1 Okt. 2019
  • EasyStat

    Dear Hanna,

    I had sent the printouts in a separate mail 5h ago, did you get it?

    The SN is 1117031005

    Best regards 
    Tobias


    durch Tobias Pfeifer - 05:26 - 1 Okt. 2019
  • RE: EasyStat BD

    Dear Tobias,

     

    Thank you for your email.

     

    1. Could you please let us know, what is the serial number on this ES analyzer?

    2. Please send us all the print outs that you have (such as calibration print out and ets. )

     

    Looking forward to hearing back from you!

     

    Best regards,

     

     

    Hanna Parhizkar | Technical Support  | hparhizkar@medicacorp.com

    Medica Corporation | 5 Oak Park Drive, Bedford, MA 01730 | www.medicacorp.com

     

    +1 781 275 4892 Main

    +1 781 541 7403 Direct                    

    +1 781 275 2731 Fax

     

     

    Signature031319_96dpi400px-01

     

     

     

     

     

    From: Tobias Pfeifer <tobias.pfeifer@fameco.eu>
    Sent: Tuesday, October 1, 2019 11:47 AM
    To: MEDICA TechnicalSupport <medicatechnicalsupport@fameco.odoo.com>
    Subject: EasyStat BD

     

    Dear Hanna,
    Dear MEDICA,

    we have an EasyStat thats acting up - it was fine for 1 month, then calibration failed overnight with BD failure.

    We had  80/24 BD Test fail

    performed cleaning with warm water

    produced 97/24 BD Test pass

    performed calibration failed 82/24 BD fail

    BD Test showed 80/24

    FLOW Tests Calibr A B C & Sample Test all ok

    where is the culprit?

    looking for some guidance on troubleshooting since the clinic is filled with patients tonight

    best regards
    Tobias


    --

    Image removed by sender.

    Tobias Pfeifer | Geschäftsführung

    T: +49 89 248805810  | F: +49 89 248805819

    E: tobias.pfeifer@fameco.eu | W: https://fameco.eu/

    FAMECO GmbH | Geschäftsführer: Tobias Pfeifer, Thomas Andersen

    Neumarkter Str. 34, 81673 München

    USt-IdNr. DE323800861 | Amtsgericht München HRB 248816

    The company accepts no liability for the content of this email, or for the consequences of any actions taken on the basis of the information provided, unless that information is subsequently confirmed in writing. If you are not the intended recipient you are notified that disclosing, copying, distributing or taking any action in reliance on the contents of this information is strictly prohibited.

     

    Sent by FAMECO GmbH using Odoo.


    durch Hanna Parhizkar - 03:36 - 1 Okt. 2019
  • EasyStat BD

    Attached are the printouts from our troubleshooting


    durch Tobias Pfeifer - 11:03 - 1 Okt. 2019