MEDICA TechnicalSupport
medicatechnicalsupport@fameco.odoo.com
-
RE: EasyStat Na Hct
Dear Tobias,
Thank you very much for the print outs.
Based on the print outs the analyzer is working well.
The percentage of red blood cells to the total blood volume is defined as hematocrit. It is quantified by measuring the electrical impedance of blood sample. Two standard solutions are used to calibrate the hematocrit sensor. The analyzer then measures the electrical impedance of the blood sample to obtain the hematocrit value. Next, the hematocrit value obtained is corrected for the concentration of the sodium ion.( Please see the attached file from user manual)
So you always need to have a working sodium electrode ( with the good calibration results and QC results ) to have the good results for HCT.
Sorry for the confusion Tobias.
Best Regards,
Hanna Parhizkar | Technical Support | hparhizkar@medicacorp.com
Medica Corporation | 5 Oak Park Drive, Bedford, MA 01730 | www.medicacorp.com
+1 781 275 4892 Main
+1 781 541 7403 Direct
+1 781 275 2731 Fax
From: Tobias Pfeifer <tobias.pfeifer@fameco.eu>
Sent: Tuesday, October 22, 2019 4:14 AM
To: MEDICA TechnicalSupport <medicatechnicalsupport@fameco.odoo.com>
Subject: EasyStat Na HctDear Hanna,
Here are the printouts I got from the Na Hct problem.
Best regards
Tobias
--Tobias Pfeifer | Geschäftsführung
T: +49 89 248805810 | F: +49 89 248805819
FAMECO GmbH | Geschäftsführer: Tobias Pfeifer, Thomas Andersen
Neumarkter Str. 34, 81673 München
USt-IdNr. DE323800861 | Amtsgericht München HRB 248816
The company accepts no liability for the content of this email, or for the consequences of any actions taken on the basis of the information provided, unless that information is subsequently confirmed in writing. If you are not the intended recipient you are notified that disclosing, copying, distributing or taking any action in reliance on the contents of this information is strictly prohibited.
Sent by FAMECO GmbH using Odoo.
by Hanna Parhizkar - 02:56 - 22 Oct 2019 -
EasyStat Na Hct
Dear Hanna,
Here are the printouts I got from the Na Hct problem.
Best regards
Tobias
by Tobias Pfeifer - 04:13 - 22 Oct 2019 -
RE: ES Na & Hct
Dear Tobias,
Thank you for your email.
- Could you please let us know what is the serial number on this analyzer?
- Could you please send us the calibration print out?
- You mentioned that whenever the Na is not working the Hct is not working either. Could you please send us a print out?
- Which kind of sample do they test?
- Which kind of sample tubing do they use?
- What is the part number for the daily cleaner? For how long do they use this fresh daily cleaner after making a new one?
- No, Na and Hct don’t have any connection.
Looking forward to hearing back from you!
Best regards,
Hanna Parhizkar | Technical Support | hparhizkar@medicacorp.com
Medica Corporation | 5 Oak Park Drive, Bedford, MA 01730 | www.medicacorp.com
+1 781 275 4892 Main
+1 781 541 7403 Direct
+1 781 275 2731 Fax
From: Tobias Pfeifer <tobias.pfeifer@fameco.eu>
Sent: Wednesday, October 16, 2019 7:53 AM
To: MEDICA TechnicalSupport <medicatechnicalsupport@fameco.odoo.com>
Subject: ES Na & HctDear Hanna,
I have a doctor who claims that whenever the Na is not working also the Hct is not working.
He needs Hct but doesnt need Na at all.
Is there a connection?
best regards
Tobias
--Tobias Pfeifer | Geschäftsführung
T: +49 89 248805810 | F: +49 89 248805819
FAMECO GmbH | Geschäftsführer: Tobias Pfeifer, Thomas Andersen
Neumarkter Str. 34, 81673 München
USt-IdNr. DE323800861 | Amtsgericht München HRB 248816
The company accepts no liability for the content of this email, or for the consequences of any actions taken on the basis of the information provided, unless that information is subsequently confirmed in writing. If you are not the intended recipient you are notified that disclosing, copying, distributing or taking any action in reliance on the contents of this information is strictly prohibited.
Sent by FAMECO GmbH using Odoo.
by Hanna Parhizkar - 04:16 - 16 Oct 2019 -
RE: ES & EBG packs policy
Dear Tobias,
The shelf life for the ES and EBG reagent modules are one year from the date of manufacture. Please make sure to use them before this time.
Es and EBG reagent modules have dallas chip. When you install these reagent modules, the dallas chip reads the date and analyzer serial number ect.
When you install the EBG reagent module you can use it for 45 days. You can use ES reagent module for 35 days when you install it.
Regarding your questions:
- No, you can only use it until the expiration date. When it is expired you can’t use it any more. If you install one reagent module on it’s last day you can’t use it the day after or for 35 or 45 days because it will give you “expired reagent module” message.
- Unfortunately we don’t have this option on our reagent modules and you can swap them between analyzers.
But you can see from the print out that this reagent module was installed on which EBG analyzer for the first time:
- Please press 5 for second menu
- Press 4 for stored information
- Press 5 for reagent module
- Press 1 for status. Here you can see the analyzer’s serial number that the reagent module was installed for the first time.
Hope it helps.
Please contact us if you have any questions.
Best regards,
Hanna Parhizkar | Technical Support | hparhizkar@medicacorp.com
Medica Corporation | 5 Oak Park Drive, Bedford, MA 01730 | www.medicacorp.com
+1 781 275 4892 Main
+1 781 541 7403 Direct
+1 781 275 2731 Fax
From: Tobias Pfeifer <tobias.pfeifer@fameco.eu>
Sent: Wednesday, October 16, 2019 7:52 AM
To: MEDICA TechnicalSupport <medicatechnicalsupport@fameco.odoo.com>
Subject: ES & EBG packs policyDear Hanna,
two questions regarding the packs:
1. will the pack if installed on the last day of it's expiry (IN date) work through the whole 45 days, or will it be cut of to the date thats on the box?
2. where can I find guidance on how the packs are protected against swapping from one EBG to another EBG for example?
best regards
Tobias
--Tobias Pfeifer | Geschäftsführung
T: +49 89 248805810 | F: +49 89 248805819
FAMECO GmbH | Geschäftsführer: Tobias Pfeifer, Thomas Andersen
Neumarkter Str. 34, 81673 München
USt-IdNr. DE323800861 | Amtsgericht München HRB 248816
The company accepts no liability for the content of this email, or for the consequences of any actions taken on the basis of the information provided, unless that information is subsequently confirmed in writing. If you are not the intended recipient you are notified that disclosing, copying, distributing or taking any action in reliance on the contents of this information is strictly prohibited.
Sent by FAMECO GmbH using Odoo.
by Hanna Parhizkar - 04:00 - 16 Oct 2019 -
ES Na & Hct
Dear Hanna,
I have a doctor who claims that whenever the Na is not working also the Hct is not working.
He needs Hct but doesnt need Na at all.
Is there a connection?
best regards
Tobias
by Tobias Pfeifer - 07:52 - 16 Oct 2019 -
ES & EBG packs policy
Dear Hanna,
two questions regarding the packs:
1. will the pack if installed on the last day of it's expiry (IN date) work through the whole 45 days, or will it be cut of to the date thats on the box?
2. where can I find guidance on how the packs are protected against swapping from one EBG to another EBG for example?
best regards
Tobias
by Tobias Pfeifer - 07:51 - 16 Oct 2019 -
RE: EBG BD
Dear Tobias,
Thank you for your email.
The top bubble detector and the bottom bubble detector are different. ( Please see the attached pictures for your information)
Depending on the error that you get you might need one of them or both of them. I would recommend to have both of them with you when you go for repair.
Please see the attached tech tip for your information.
Please contact us if you have any questions.
Best regards,
Hanna Parhizkar | Technical Support | hparhizkar@medicacorp.com
Medica Corporation | 5 Oak Park Drive, Bedford, MA 01730 | www.medicacorp.com
+1 781 275 4892 Main
+1 781 541 7403 Direct
+1 781 275 2731 Fax
From: Tobias Pfeifer <tobias.pfeifer@fameco.eu>
Sent: Tuesday, October 15, 2019 3:25 PM
To: MEDICA TechnicalSupport <medicatechnicalsupport@fameco.odoo.com>
Subject: EBG BDDear Hanna,
What's the difference between 6801 and 6802 BD Kit Top and bottom, do I need both for 1 repair?
Best regards
Tobias
--Tobias Pfeifer | Geschäftsführung
T: +49 89 248805810 | F: +49 89 248805819
FAMECO GmbH | Geschäftsführer: Tobias Pfeifer, Thomas Andersen
Neumarkter Str. 34, 81673 München
USt-IdNr. DE323800861 | Amtsgericht München HRB 248816
The company accepts no liability for the content of this email, or for the consequences of any actions taken on the basis of the information provided, unless that information is subsequently confirmed in writing. If you are not the intended recipient you are notified that disclosing, copying, distributing or taking any action in reliance on the contents of this information is strictly prohibited.
Sent by FAMECO GmbH using Odoo.
Sent from my iPhone
Sent from my iPhone
by Hanna Parhizkar - 04:51 - 15 Oct 2019 -
EBG BD
Dear Hanna,
What's the difference between 6801 and 6802 BD Kit Top and bottom, do I need both for 1 repair?
Best regards
Tobias
by Tobias Pfeifer - 03:24 - 15 Oct 2019 -
RE: Cost per Test
Dear Tobias,
I forwarded your email to the appropriate person. He is working on these questions. As soon as he sends me the answers, I will forward them to you.
Thank you for your patience.
Best regards,
Hanna Parhizkar | Technical Support | hparhizkar@medicacorp.com
Medica Corporation | 5 Oak Park Drive, Bedford, MA 01730 | www.medicacorp.com
+1 781 275 4892 Main
+1 781 541 7403 Direct
+1 781 275 2731 Fax
From: Tobias Pfeifer <tobias.pfeifer@fameco.eu>
Sent: Tuesday, October 8, 2019 2:30 PM
To: MEDICA TechnicalSupport <medicatechnicalsupport@fameco.odoo.com>
Subject: Cost per TestDear Hanna,
any news on the QC usage?
Who could I contact for some general guidance on the supplied XLS calculating worksheets?
best regards
Tobias
--Tobias Pfeifer | Geschäftsführung
T: +49 89 248805810 | F: +49 89 248805819
FAMECO GmbH | Geschäftsführer: Tobias Pfeifer, Thomas Andersen
Neumarkter Str. 34, 81673 München
USt-IdNr. DE323800861 | Amtsgericht München HRB 248816
The company accepts no liability for the content of this email, or for the consequences of any actions taken on the basis of the information provided, unless that information is subsequently confirmed in writing. If you are not the intended recipient you are notified that disclosing, copying, distributing or taking any action in reliance on the contents of this information is strictly prohibited.
Sent by FAMECO GmbH using Odoo.
by Hanna Parhizkar - 10:41 - 9 Oct 2019 -
Cost per Test
Dear Hanna,
any news on the QC usage?
Who could I contact for some general guidance on the supplied XLS calculating worksheets?
best regards
Tobias
by Tobias Pfeifer - 02:29 - 8 Oct 2019 -
RE: EasyLyte QC usage
Dear Tobias,
I will forward your email to the appropriate person. As soon as I get the answer I will forward it to you.
Best regards,
Hanna Parhizkar | Technical Support | hparhizkar@medicacorp.com
Medica Corporation | 5 Oak Park Drive, Bedford, MA 01730 | www.medicacorp.com
+1 781 275 4892 Main
+1 781 541 7403 Direct
+1 781 275 2731 Fax
From: Tobias Pfeifer <tobias.pfeifer@fameco.eu>
Sent: Monday, October 7, 2019 9:36 AM
To: MEDICA TechnicalSupport <medicatechnicalsupport@fameco.odoo.com>
Subject: EasyLyte QC usageHey,
we have a new potential customer for a EL Expand;
they are doing 600 samples a week;
they want to do QC everyday - is there a tool to calculate amounts and costs for the QCs?
Sodium/Potassium/Chloride/Lithium Analyzer
800 mL Reagent Pack Life Cost Per Test
Samples Total(mL) Pack Total Samples Reagent Total Cost/ Total Cost/
per 5day Std A Life Samples per 5day Cost per Sample * Test **
Week Usage/Week (weeks) per Pack Week Sample
135 98,8 8,1 1093 135 ,16 ,39 $0,10
100 80,6 9,9 993 100 ,17 ,49 $0,12
50 54,6 14,7 733 50 ,24 ,86 $0,22
40 49,4 16,2 648 40 ,27 1,05 $0,26
35 46,8 17,1 598 35 ,29 1,18 $0,30
30 44,2 18,1 543 30 ,32 1,36 $0,34
25 41,6 19,2 481 25 ,36 1,61 $0,40
20 39,0 20,5 410 20 ,42 1,98 $0,50
15 36,4 22,0 330 15 ,52 2,61 $0,65
Enter Samples per 5day Week* Enter Samples per 5day Week
150 106,6 7,5 1126 150 ,15 ,39 ,10
*Use for samples greater than 135 * Includes all reagents, maintenance, and supply
items for 1 year
** 1 Sample = 3 Tests, Na,K and Li
Assumptions Yearly Supply Requirements
Samples per 5day Week, less than 135
Reagent Pack Volume (mL) 800 Quantity** Description Price Total
EasyLyte Price 3970,$00 2,0 Maintenance Kits 75,$00 150,$00
Reimbursement/Sample 12,$81 0,7 Na Electrode 213,00 149,10
Reagent Pack Cost 173,$00 1,3 K Electrode 190,00 247,00
1,3 Li Electrode 250,00 325,00
Standard A Usage (mL) 1,3 Cl Electrode 190,00 247,00
Sample Analysis ,52 4,0 Disp. Ref. Electrode 95,00 380,00
1st Calibration/day 1,58 2,0 Daily Cleaning Solution 40,00 80,00
2nd, 3rd, 4th Cals ,50 2,0 Packages Printer Paper 23,00 46,00
Sip ,04 1624,$10
Cleaner ,32 ** Electrode usage based on typical useful electrode life
Yearly Supply Requirements
Samples per 5day Week, less than 135
Base Standard A Usage Quantity*** Description Price Total
7 Calibrations/week 13,56 2,0 Maintenance Kits 75,$00 150,$00
5 Cleanings/week 1,60 0,8 Na Electrode 213,00 170,40
336 Sips per week 13,44 1,5 K Electrode 190,00 285,00
Total 28,60 1,5 Li Electrode 250,00 375,00
1,5 Cl Electrode 190,00 285,00
4,0 Disp. Ref. Electrode 95,00 380,00
3,0 Daily Cleaning Solution 40,00 120,00
3,0 Packages Printer Paper 23,00 69,00
1834,$40
***Calculations derived from manual entry of sample volume
best regards
Tobias
--Tobias Pfeifer | Geschäftsführung
T: +49 89 248805810 | F: +49 89 248805819
FAMECO GmbH | Geschäftsführer: Tobias Pfeifer, Thomas Andersen
Neumarkter Str. 34, 81673 München
USt-IdNr. DE323800861 | Amtsgericht München HRB 248816
The company accepts no liability for the content of this email, or for the consequences of any actions taken on the basis of the information provided, unless that information is subsequently confirmed in writing. If you are not the intended recipient you are notified that disclosing, copying, distributing or taking any action in reliance on the contents of this information is strictly prohibited.
Sent by FAMECO GmbH using Odoo.
by Hanna Parhizkar - 02:27 - 7 Oct 2019 -
EasyLyte QC usage
Hey,
we have a new potential customer for a EL Expand;
they are doing 600 samples a week;
they want to do QC everyday - is there a tool to calculate amounts and costs for the QCs?
Sodium/Potassium/Chloride/Lithium Analyzer
800 mL Reagent Pack Life Cost Per Test
Samples Total(mL) Pack Total Samples Reagent Total Cost/ Total Cost/
per 5day Std A Life Samples per 5day Cost per Sample * Test **
Week Usage/Week (weeks) per Pack Week Sample
135 98,8 8,1 1093 135 ,16 ,39 $0,10
100 80,6 9,9 993 100 ,17 ,49 $0,12
50 54,6 14,7 733 50 ,24 ,86 $0,22
40 49,4 16,2 648 40 ,27 1,05 $0,26
35 46,8 17,1 598 35 ,29 1,18 $0,30
30 44,2 18,1 543 30 ,32 1,36 $0,34
25 41,6 19,2 481 25 ,36 1,61 $0,40
20 39,0 20,5 410 20 ,42 1,98 $0,50
15 36,4 22,0 330 15 ,52 2,61 $0,65
Enter Samples per 5day Week* Enter Samples per 5day Week
150 106,6 7,5 1126 150 ,15 ,39 ,10
*Use for samples greater than 135 * Includes all reagents, maintenance, and supply
items for 1 year
** 1 Sample = 3 Tests, Na,K and Li
Assumptions Yearly Supply Requirements
Samples per 5day Week, less than 135
Reagent Pack Volume (mL) 800 Quantity** Description Price Total
EasyLyte Price 3970,$00 2,0 Maintenance Kits 75,$00 150,$00
Reimbursement/Sample 12,$81 0,7 Na Electrode 213,00 149,10
Reagent Pack Cost 173,$00 1,3 K Electrode 190,00 247,00
1,3 Li Electrode 250,00 325,00
Standard A Usage (mL) 1,3 Cl Electrode 190,00 247,00
Sample Analysis ,52 4,0 Disp. Ref. Electrode 95,00 380,00
1st Calibration/day 1,58 2,0 Daily Cleaning Solution 40,00 80,00
2nd, 3rd, 4th Cals ,50 2,0 Packages Printer Paper 23,00 46,00
Sip ,04 1624,$10
Cleaner ,32 ** Electrode usage based on typical useful electrode life
Yearly Supply Requirements
Samples per 5day Week, less than 135
Base Standard A Usage Quantity*** Description Price Total
7 Calibrations/week 13,56 2,0 Maintenance Kits 75,$00 150,$00
5 Cleanings/week 1,60 0,8 Na Electrode 213,00 170,40
336 Sips per week 13,44 1,5 K Electrode 190,00 285,00
Total 28,60 1,5 Li Electrode 250,00 375,00
1,5 Cl Electrode 190,00 285,00
4,0 Disp. Ref. Electrode 95,00 380,00
3,0 Daily Cleaning Solution 40,00 120,00
3,0 Packages Printer Paper 23,00 69,00
1834,$40
***Calculations derived from manual entry of sample volume
best regards
Tobias
by Tobias Pfeifer - 09:35 - 7 Oct 2019 -
RE: EasyLyte -2- claims
Dear Tobias,
Thank you for the failure report.
On the EL EXPAND analyzer, every time that you have shifts in QC or high slope failure ( Calibration value ), please change the reference electrode and send us the reference electrode lot number, calibration results, QC data. We will send you a replacement for the reference electrode and make an RGA for you.
We would recommend to run the daily cleaner in the morning, do calibration and run the QC materials. If they are good and within the normal ranges, run the patient samples.
I processed the replacements for you. If you have any questions or concerns regarding the shipment of this order, please contact our Customer Service directly at : CustomerService@medicacorp.com
We would like to get the reference electrode back to Medica for further investigations. I made the RGA 192085 for you. Please use our FedEx account number 471164747 and forward it to the address below:
Medica Corporation
Attn: Hanna Parhizkar ( RGA 192085 )
3 Oak Park Drive
Bedford, MA 01730
Tel: 781-275-4892
Please make sure this RGA number is visible on the external shipping crate. Please send us the tracking number when it is ready.
Best regards,
Hanna Parhizkar | Technical Support | hparhizkar@medicacorp.com
Medica Corporation | 5 Oak Park Drive, Bedford, MA 01730 | www.medicacorp.com
+1 781 275 4892 Main
+1 781 541 7403 Direct
+1 781 275 2731 Fax
From: Tobias Pfeifer <tobias.pfeifer@fameco.eu>
Sent: Wednesday, October 2, 2019 2:38 AM
To: MEDICA TechnicalSupport <medicatechnicalsupport@fameco.odoo.com>
Subject: EasyLyte -2- claimsDear Hanna,
here are 2 claims for EasyLyte Electrodes after we finished troubleshooting and exchanging parts.
best regards
Tobias
--Tobias Pfeifer | Geschäftsführung
T: +49 89 248805810 | F: +49 89 248805819
FAMECO GmbH | Geschäftsführer: Tobias Pfeifer, Thomas Andersen
Neumarkter Str. 34, 81673 München
USt-IdNr. DE323800861 | Amtsgericht München HRB 248816
The company accepts no liability for the content of this email, or for the consequences of any actions taken on the basis of the information provided, unless that information is subsequently confirmed in writing. If you are not the intended recipient you are notified that disclosing, copying, distributing or taking any action in reliance on the contents of this information is strictly prohibited.
Sent by FAMECO GmbH using Odoo.
Dear Tobias,
Your warranty order #TSW-192085 has been processed. Attached is a summary of your order. You will receive a formal order acknowledgement from our Customer Service Team shortly.
Best Regards,
Hanna Parhizkar
by Hanna Parhizkar - 02:16 - 2 Oct 2019 -
EasyStat BD & Calibration
Dear Hanna,
please see this most recent printouts - how come BD Test is good - Calibration than bad, then good - it's working as of this moment but nobody trusts the instrument anymore until we can find some explanation and final solution.
best regards
Tobias
by Tobias Pfeifer - 05:56 - 2 Oct 2019 -
EasyLyte -2- claims
Dear Hanna,
here are 2 claims for EasyLyte Electrodes after we finished troubleshooting and exchanging parts.
best regards
Tobias
by Tobias Pfeifer - 02:38 - 2 Oct 2019 -
EasysStat SN 1117031005
Dear Hanna,
here are the printouts - after cleaning the BD and the valve module also with bleach
best regards
Tobias
by Tobias Pfeifer - 02:13 - 2 Oct 2019 -
Re: EasyStat
Dear Tobias,
Unfortunately I didn’t get the print outs.
Could you please re-send the print outs?
Thank you very much!
Best Regards,
Hanna
From: Tobias Pfeifer <tobias.pfeifer@fameco.eu>
Sent: Tuesday, October 1, 2019 6:26:49 PM
To: MEDICA TechnicalSupport
Subject: EasyStatDear Hanna,
I had sent the printouts in a separate mail 5h ago, did you get it?
The SN is 1117031005
Best regards
Tobias
--
Tobias Pfeifer | Geschäftsführung
T: +49 89 248805810 | F: +49 89 248805819
FAMECO GmbH | Geschäftsführer: Tobias Pfeifer, Thomas Andersen
Neumarkter Str. 34, 81673 München
USt-IdNr. DE323800861 | Amtsgericht München HRB 248816
The company accepts no liability for the content of this email, or for the consequences of any actions taken on the basis of the information provided, unless that information is subsequently confirmed in writing. If you are not the intended recipient you are notified that disclosing, copying, distributing or taking any action in reliance on the contents of this information is strictly prohibited.
Sent by FAMECO GmbH using Odoo.
by Hanna Parhizkar - 08:58 - 1 Oct 2019 -
EasyStat
Dear Hanna,
I had sent the printouts in a separate mail 5h ago, did you get it?
The SN is 1117031005
Best regards
Tobias
by Tobias Pfeifer - 06:26 - 1 Oct 2019 -
RE: EasyStat BD
Dear Tobias,
Thank you for your email.
- Could you please let us know, what is the serial number on this ES analyzer?
- Please send us all the print outs that you have (such as calibration print out and ets. )
Looking forward to hearing back from you!
Best regards,
Hanna Parhizkar | Technical Support | hparhizkar@medicacorp.com
Medica Corporation | 5 Oak Park Drive, Bedford, MA 01730 | www.medicacorp.com
+1 781 275 4892 Main
+1 781 541 7403 Direct
+1 781 275 2731 Fax
From: Tobias Pfeifer <tobias.pfeifer@fameco.eu>
Sent: Tuesday, October 1, 2019 11:47 AM
To: MEDICA TechnicalSupport <medicatechnicalsupport@fameco.odoo.com>
Subject: EasyStat BDDear Hanna,
Dear MEDICA,
we have an EasyStat thats acting up - it was fine for 1 month, then calibration failed overnight with BD failure.
We had 80/24 BD Test fail
performed cleaning with warm water
produced 97/24 BD Test pass
performed calibration failed 82/24 BD fail
BD Test showed 80/24
FLOW Tests Calibr A B C & Sample Test all ok
where is the culprit?
looking for some guidance on troubleshooting since the clinic is filled with patients tonight
best regards
Tobias
--Tobias Pfeifer | Geschäftsführung
T: +49 89 248805810 | F: +49 89 248805819
FAMECO GmbH | Geschäftsführer: Tobias Pfeifer, Thomas Andersen
Neumarkter Str. 34, 81673 München
USt-IdNr. DE323800861 | Amtsgericht München HRB 248816
The company accepts no liability for the content of this email, or for the consequences of any actions taken on the basis of the information provided, unless that information is subsequently confirmed in writing. If you are not the intended recipient you are notified that disclosing, copying, distributing or taking any action in reliance on the contents of this information is strictly prohibited.
Sent by FAMECO GmbH using Odoo.
by Hanna Parhizkar - 04:36 - 1 Oct 2019 -
EasyStat BD
Attached are the printouts from our troubleshooting
by Tobias Pfeifer - 12:03 - 1 Oct 2019