Dear Tobias,
Thank you very much for the print outs.
Based on the print outs the analyzer is working well.
The percentage of red blood cells to the total blood volume is defined as hematocrit. It is quantified by measuring the electrical impedance of blood sample. Two standard solutions are used to calibrate the hematocrit sensor. The analyzer then measures the electrical impedance of the blood sample to obtain the hematocrit value. Next, the hematocrit value obtained is corrected for the concentration of the sodium ion.( Please see the attached file from user manual)
So you always need to have a working sodium electrode ( with the good calibration results and QC results ) to have the good results for HCT.
Sorry for the confusion Tobias.
Best Regards,
Hanna Parhizkar | Technical Support | hparhizkar@medicacorp.com
Medica Corporation | 5 Oak Park Drive, Bedford, MA 01730 | www.medicacorp.com
+1 781 275 4892 Main
+1 781 541 7403 Direct
+1 781 275 2731 Fax
From: Tobias Pfeifer <tobias.pfeifer@fameco.eu>
Sent: Tuesday, October 22, 2019 4:14 AM
To: MEDICA TechnicalSupport <medicatechnicalsupport@fameco.odoo.com>
Subject: EasyStat Na Hct
Dear Hanna,
Here are the printouts I got from the Na Hct problem.
Best regards
Tobias
--
Tobias Pfeifer | Geschäftsführung |
T: +49 89 248805810 | F: +49 89 248805819 |
FAMECO GmbH | Geschäftsführer: Tobias Pfeifer, Thomas Andersen |
Neumarkter Str. 34, 81673 München |
USt-IdNr. DE323800861 | Amtsgericht München HRB 248816 |
The company accepts no liability for the content of this email, or for the consequences of any actions taken on the basis of the information provided, unless that information is subsequently confirmed in writing. If you are not the intended recipient you are notified that disclosing, copying, distributing or taking any action in reliance on the contents of this information is strictly prohibited. |
Sent by FAMECO GmbH using Odoo.
durch Hanna Parhizkar - 02:56 - 22 Okt. 2019
Referenz
-
EasyStat Na Hct
Dear Hanna,
Here are the printouts I got from the Na Hct problem.
Best regards
Tobias
durch Tobias Pfeifer - 04:13 - 22 Okt. 2019-
RE: EasyStat Na Hct
Dear Tobias,
Thank you very much for the print outs.
Based on the print outs the analyzer is working well.
The percentage of red blood cells to the total blood volume is defined as hematocrit. It is quantified by measuring the electrical impedance of blood sample. Two standard solutions are used to calibrate the hematocrit sensor. The analyzer then measures the electrical impedance of the blood sample to obtain the hematocrit value. Next, the hematocrit value obtained is corrected for the concentration of the sodium ion.( Please see the attached file from user manual)
So you always need to have a working sodium electrode ( with the good calibration results and QC results ) to have the good results for HCT.
Sorry for the confusion Tobias.
Best Regards,
Hanna Parhizkar | Technical Support | hparhizkar@medicacorp.com
Medica Corporation | 5 Oak Park Drive, Bedford, MA 01730 | www.medicacorp.com
+1 781 275 4892 Main
+1 781 541 7403 Direct
+1 781 275 2731 Fax
From: Tobias Pfeifer <tobias.pfeifer@fameco.eu>
Sent: Tuesday, October 22, 2019 4:14 AM
To: MEDICA TechnicalSupport <medicatechnicalsupport@fameco.odoo.com>
Subject: EasyStat Na HctDear Hanna,
Here are the printouts I got from the Na Hct problem.
Best regards
Tobias
--Tobias Pfeifer | Geschäftsführung
T: +49 89 248805810 | F: +49 89 248805819
FAMECO GmbH | Geschäftsführer: Tobias Pfeifer, Thomas Andersen
Neumarkter Str. 34, 81673 München
USt-IdNr. DE323800861 | Amtsgericht München HRB 248816
The company accepts no liability for the content of this email, or for the consequences of any actions taken on the basis of the information provided, unless that information is subsequently confirmed in writing. If you are not the intended recipient you are notified that disclosing, copying, distributing or taking any action in reliance on the contents of this information is strictly prohibited.
Sent by FAMECO GmbH using Odoo.
durch Hanna Parhizkar - 02:56 - 22 Okt. 2019
-