- Mailinglisten
- MEDICA TechnicalSupport
- RE: EasyStat BD
Archive
MEDICA TechnicalSupport
medicatechnicalsupport@fameco.odoo.com
RE: EasyStat BD
Dear Tobias,
Thank you for your email.
- Could you please let us know, what is the serial number on this ES analyzer?
- Please send us all the print outs that you have (such as calibration print out and ets. )
Looking forward to hearing back from you!
Best regards,
Hanna Parhizkar | Technical Support | hparhizkar@medicacorp.com
Medica Corporation | 5 Oak Park Drive, Bedford, MA 01730 | www.medicacorp.com
+1 781 275 4892 Main
+1 781 541 7403 Direct
+1 781 275 2731 Fax
From: Tobias Pfeifer <tobias.pfeifer@fameco.eu>
Sent: Tuesday, October 1, 2019 11:47 AM
To: MEDICA TechnicalSupport <medicatechnicalsupport@fameco.odoo.com>
Subject: EasyStat BD
Dear Hanna,
Dear MEDICA,
we have an EasyStat thats acting up - it was fine for 1 month, then calibration failed overnight with BD failure.
We had 80/24 BD Test fail
performed cleaning with warm water
produced 97/24 BD Test pass
performed calibration failed 82/24 BD fail
BD Test showed 80/24
FLOW Tests Calibr A B C & Sample Test all ok
where is the culprit?
looking for some guidance on troubleshooting since the clinic is filled with patients tonight
best regards
Tobias
--
Tobias Pfeifer | Geschäftsführung |
T: +49 89 248805810 | F: +49 89 248805819 |
FAMECO GmbH | Geschäftsführer: Tobias Pfeifer, Thomas Andersen |
Neumarkter Str. 34, 81673 München |
USt-IdNr. DE323800861 | Amtsgericht München HRB 248816 |
The company accepts no liability for the content of this email, or for the consequences of any actions taken on the basis of the information provided, unless that information is subsequently confirmed in writing. If you are not the intended recipient you are notified that disclosing, copying, distributing or taking any action in reliance on the contents of this information is strictly prohibited. |
Sent by FAMECO GmbH using Odoo.
durch Hanna Parhizkar - 04:36 - 1 Okt. 2019
Referenz
-
EasyStat BD
Dear Hanna,
Dear MEDICA,
we have an EasyStat thats acting up - it was fine for 1 month, then calibration failed overnight with BD failure.
We had 80/24 BD Test fail
performed cleaning with warm water
produced 97/24 BD Test pass
performed calibration failed 82/24 BD fail
BD Test showed 80/24
FLOW Tests Calibr A B C & Sample Test all ok
where is the culprit?
looking for some guidance on troubleshooting since the clinic is filled with patients tonight
best regards
Tobias
durch Tobias Pfeifer - 11:46 - 1 Okt. 2019