Dear Hanna,
a clinic has just installed a new reagentmodule 19071029 - they have now faced a variety of problems with pH noise and pO2 drift.
See attached printouts. Should we troubleshoot the module, or the device?
best regards
Tobias
durch Tobias Pfeifer - 06:20 - 5 Nov. 2019
Folgenachrichten
-
RE: ES noise & drift
Dear Tobias,
Thank you for the information and print outs.
The ES analyzer will not show the calculated results if the calibration results are not good or if the calculated results are out of the range. ( Please see the attached from user manual )
- Looking at the calibration print outs they are all good and within the normal ranges.
- Could you please let us know what is the serial number on this analyzer?
- Are you using Medica capillary tubing kit for ES? Please send us the part number or a picture of the kit.
- Which kind of sample are you testing?
- The print outs show that you are using the capillary mode. In the capillary mode the chance of the pH getting exposed to air is very high. When pH changes the PCO2 and PO2 also change.
- Please run the normal QC as a patient sample and see if you get the same error. Please send us the calibration print out and this print out.
Best Regards,
Hanna Parhizkar | Technical Support | hparhizkar@medicacorp.com
Medica Corporation | 5 Oak Park Drive, Bedford, MA 01730 | www.medicacorp.com
+1 781 275 4892 Main
+1 781 541 7403 Direct
+1 781 275 2731 Fax
From: Tobias Pfeifer <tobias.pfeifer@fameco.eu>
Sent: Tuesday, November 5, 2019 6:20 AM
To: MEDICA TechnicalSupport <medicatechnicalsupport@fameco.odoo.com>
Subject: ES noise & driftDear Hanna,
a clinic has just installed a new reagentmodule 19071029 - they have now faced a variety of problems with pH noise and pO2 drift.
See attached printouts. Should we troubleshoot the module, or the device?
best regards
Tobias
--Tobias Pfeifer | Geschäftsführung
T: +49 89 248805810 | F: +49 89 248805819
FAMECO GmbH | Geschäftsführer: Tobias Pfeifer, Thomas Andersen
Neumarkter Str. 34, 81673 München
USt-IdNr. DE323800861 | Amtsgericht München HRB 248816
The company accepts no liability for the content of this email, or for the consequences of any actions taken on the basis of the information provided, unless that information is subsequently confirmed in writing. If you are not the intended recipient you are notified that disclosing, copying, distributing or taking any action in reliance on the contents of this information is strictly prohibited.
Sent by FAMECO GmbH using Odoo.
durch Hanna Parhizkar - 02:42 - 5 Nov. 2019