RE: EasyLyte Spare Part
Dear Tobias,
Unfortunately we don’t sell the probe motor separately. We sell the probe motor with the parts in the attached picture.
The part number for this ( Attached picture ) is 001392-001 ASSY,BRACKET, MOTOR,LEAD SCREW.
Have a nice weekend.
Best regards,
Hanna Parhizkar | Technical Support | hparhizkar@medicacorp.com
Medica Corporation | 5 Oak Park Drive, Bedford, MA 01730 | www.medicacorp.com
+1 781 275 4892 Main
+1 781 541 7403 Direct
+1 781 275 2731 Fax
From: Tobias Pfeifer <tobias.pfeifer@fameco.eu>
Sent: Friday, December 6, 2019 7:12 AM
To: MEDICA TechnicalSupport <medicatechnicalsupport@fameco.odoo.com>
Subject: EasyLyte Spare PartDear Hanna,
I am looking for the part number of the attached motor
best regards
Tobias
--Tobias Pfeifer | Geschäftsführung
T: +49 89 248805810 | F: +49 89 248805819
FAMECO GmbH | Geschäftsführer: Tobias Pfeifer, Thomas Andersen
Neumarkter Str. 34, 81673 München
USt-IdNr. DE323800861 | Amtsgericht München HRB 248816
The company accepts no liability for the content of this email, or for the consequences of any actions taken on the basis of the information provided, unless that information is subsequently confirmed in writing. If you are not the intended recipient you are notified that disclosing, copying, distributing or taking any action in reliance on the contents of this information is strictly prohibited.
Sent by FAMECO GmbH using Odoo.
Sent from my iPhone
durch Hanna Parhizkar - 03:20 - 6 Dez. 2019 -
EasyLyte Spare Part
Dear Hanna,
I am looking for the part number of the attached motor
best regards
Tobias
durch Tobias Pfeifer - 07:11 - 6 Dez. 2019 -
RE: ES ph noise 1214031005
Dear Tobias,
Thank you for the update.
How about changing the sampler assembly? Did they try it?
Please let me know if changing the electrodes helped.
Best regards,
Hanna Parhizkar | Technical Support | hparhizkar@medicacorp.com
Medica Corporation | 5 Oak Park Drive, Bedford, MA 01730 | www.medicacorp.com
+1 781 275 4892 Main
+1 781 541 7403 Direct
+1 781 275 2731 Fax
From: FAMECO \ Tobias PFEIFER <tobias.pfeifer@fameco.eu>
Sent: Monday, December 2, 2019 10:46 AM
To: MEDICA TechnicalSupport <medicatechnicalsupport@fameco.odoo.com>
Subject: Re: ES ph noise 1214031005Dear Hanna,
The probe wiper is “new” changed on 10 NOV
The problem was persistent with two ES instruments, I guess the customer wants to try and exchange the electrodes next
Tobias Pfeifer
--
Tobias Pfeifer | Geschäftsführung
T: +49 89 248805810 | F: +49 89 248805819FAMECO GmbH | Geschäftsführer: Tobias Pfeifer, Thomas Andersen
Neumarkter Str. 34, 81673 München
USt-IdNr. DE323800861 | Amtsgericht München HRB 248816
The company accepts no liability for the content of this email, or for the consequences of any actions taken on the basis of the information provided, unless that information is subsequently confirmed in writing. If you are not the intended recipient you are notified that disclosing, copying, distributing or taking any action in reliance on the contents of this information is strictly prohibited.
Am 02.12.2019 um 16:09 schrieb Hanna Parhizkar <HParhizkar@medicacorp.com>:
Dear Tobias,
Thank you for your email. I apologize for the delay in responding. Medica was closed for Thanksgiving holiday on Thursday and Friday.
- When was the last time that you changed the probe wiper? Could you please change the probe wiper?
- After changing the probe wiper if you still get the error, please change the sampler assembly .
Please let us know how it goes.
Hanna Parhizkar | Technical Support | hparhizkar@medicacorp.com
Medica Corporation | 5 Oak Park Drive, Bedford, MA 01730 | www.medicacorp.com
+1 781 275 4892 Main
+1 781 541 7403 Direct
+1 781 275 2731 Fax
From: Tobias Pfeifer <tobias.pfeifer@fameco.eu>
Sent: Wednesday, November 27, 2019 11:49 AM
To: MEDICA TechnicalSupport <medicatechnicalsupport@fameco.odoo.com>
Subject: ES ph noise 1214031005Dear Hanna,
basically the problem persists, the head of the lab told me she is performing bloodgas analysis and sample handling for the last 29 years without issure.
They are using 100 ul with 100 I.U./mL Heparin capillaries.
The patient's blood is drawn directly next to the instrument
Mainly they the problem is that the first try is a failure and consistently the second try is good
attached are some printouts
best regards
Tobias
--Tobias Pfeifer | Geschäftsführung
T: +49 89 248805810 | F: +49 89 248805819
FAMECO GmbH | Geschäftsführer: Tobias Pfeifer, Thomas Andersen
Neumarkter Str. 34, 81673 München
USt-IdNr. DE323800861 | Amtsgericht München HRB 248816
The company accepts no liability for the content of this email, or for the consequences of any actions taken on the basis of the information provided, unless that information is subsequently confirmed in writing. If you are not the intended recipient you are notified that disclosing, copying, distributing or taking any action in reliance on the contents of this information is strictly prohibited.
Sent by FAMECO GmbH using Odoo.
--
Hanna ParhizkarSent by FAMECO GmbH using Odoo.
<Changing the sampler.pdf>
--Tobias Pfeifer | Geschäftsführung
T: +49 89 248805810 | F: +49 89 248805819
FAMECO GmbH | Geschäftsführer: Tobias Pfeifer, Thomas Andersen
Neumarkter Str. 34, 81673 München
USt-IdNr. DE323800861 | Amtsgericht München HRB 248816
The company accepts no liability for the content of this email, or for the consequences of any actions taken on the basis of the information provided, unless that information is subsequently confirmed in writing. If you are not the intended recipient you are notified that disclosing, copying, distributing or taking any action in reliance on the contents of this information is strictly prohibited.
Sent by FAMECO GmbH using Odoo.
durch Hanna Parhizkar - 02:54 - 2 Dez. 2019 -
Re: ES ph noise 1214031005
Dear Hanna,
The probe wiper is “new” changed on 10 NOV
The problem was persistent with two ES instruments, I guess the customer wants to try and exchange the electrodes next
Tobias Pfeifer
--
Tobias Pfeifer | Geschäftsführung
T: +49 89 248805810 | F: +49 89 248805819
FAMECO GmbH | Geschäftsführer: Tobias Pfeifer, Thomas Andersen
Neumarkter Str. 34, 81673 München
USt-IdNr. DE323800861 | Amtsgericht München HRB 248816
The company accepts no liability for the content of this email, or for the consequences of any actions taken on the basis of the information provided, unless that information is subsequently confirmed in writing. If you are not the intended recipient you are notified that disclosing, copying, distributing or taking any action in reliance on the contents of this information is strictly prohibited.
Am 02.12.2019 um 16:09 schrieb Hanna Parhizkar <HParhizkar@medicacorp.com>:
Dear Tobias,
Thank you for your email. I apologize for the delay in responding. Medica was closed for Thanksgiving holiday on Thursday and Friday.
- When was the last time that you changed the probe wiper? Could you please change the probe wiper?
- After changing the probe wiper if you still get the error, please change the sampler assembly .
Please let us know how it goes.
Hanna Parhizkar | Technical Support | hparhizkar@medicacorp.com
Medica Corporation | 5 Oak Park Drive, Bedford, MA 01730 | www.medicacorp.com
+1 781 275 4892 Main
+1 781 541 7403 Direct
+1 781 275 2731 Fax
From: Tobias Pfeifer <tobias.pfeifer@fameco.eu>
Sent: Wednesday, November 27, 2019 11:49 AM
To: MEDICA TechnicalSupport <medicatechnicalsupport@fameco.odoo.com>
Subject: ES ph noise 1214031005Dear Hanna,
basically the problem persists, the head of the lab told me she is performing bloodgas analysis and sample handling for the last 29 years without issure.
They are using 100 ul with 100 I.U./mL Heparin capillaries.
The patient's blood is drawn directly next to the instrument
Mainly they the problem is that the first try is a failure and consistently the second try is good
attached are some printouts
best regards
Tobias
--Tobias Pfeifer | Geschäftsführung
T: +49 89 248805810 | F: +49 89 248805819
FAMECO GmbH | Geschäftsführer: Tobias Pfeifer, Thomas Andersen
Neumarkter Str. 34, 81673 München
USt-IdNr. DE323800861 | Amtsgericht München HRB 248816
The company accepts no liability for the content of this email, or for the consequences of any actions taken on the basis of the information provided, unless that information is subsequently confirmed in writing. If you are not the intended recipient you are notified that disclosing, copying, distributing or taking any action in reliance on the contents of this information is strictly prohibited.
Sent by FAMECO GmbH using Odoo.
--
Hanna ParhizkarSent by FAMECO GmbH using Odoo.
<Changing the sampler.pdf>
durch Tobias Pfeifer - 10:45 - 2 Dez. 2019 -
RE: ES ph noise 1214031005
Dear Tobias,
Thank you for your email. I apologize for the delay in responding. Medica was closed for Thanksgiving holiday on Thursday and Friday.
- When was the last time that you changed the probe wiper? Could you please change the probe wiper?
- After changing the probe wiper if you still get the error, please change the sampler assembly .
Please let us know how it goes.
Hanna Parhizkar | Technical Support | hparhizkar@medicacorp.com
Medica Corporation | 5 Oak Park Drive, Bedford, MA 01730 | www.medicacorp.com
+1 781 275 4892 Main
+1 781 541 7403 Direct
+1 781 275 2731 Fax
From: Tobias Pfeifer <tobias.pfeifer@fameco.eu>
Sent: Wednesday, November 27, 2019 11:49 AM
To: MEDICA TechnicalSupport <medicatechnicalsupport@fameco.odoo.com>
Subject: ES ph noise 1214031005Dear Hanna,
basically the problem persists, the head of the lab told me she is performing bloodgas analysis and sample handling for the last 29 years without issure.
They are using 100 ul with 100 I.U./mL Heparin capillaries.
The patient's blood is drawn directly next to the instrument
Mainly they the problem is that the first try is a failure and consistently the second try is good
attached are some printouts
best regards
Tobias
--Tobias Pfeifer | Geschäftsführung
T: +49 89 248805810 | F: +49 89 248805819
FAMECO GmbH | Geschäftsführer: Tobias Pfeifer, Thomas Andersen
Neumarkter Str. 34, 81673 München
USt-IdNr. DE323800861 | Amtsgericht München HRB 248816
The company accepts no liability for the content of this email, or for the consequences of any actions taken on the basis of the information provided, unless that information is subsequently confirmed in writing. If you are not the intended recipient you are notified that disclosing, copying, distributing or taking any action in reliance on the contents of this information is strictly prohibited.
Sent by FAMECO GmbH using Odoo.
durch Hanna Parhizkar - 10:08 - 2 Dez. 2019 -
ES ph noise 1214031005
Dear Hanna,
basically the problem persists, the head of the lab told me she is performing bloodgas analysis and sample handling for the last 29 years without issure.
They are using 100 ul with 100 I.U./mL Heparin capillaries.
The patient's blood is drawn directly next to the instrument
Mainly they the problem is that the first try is a failure and consistently the second try is good
attached are some printouts
best regards
Tobias
durch Tobias Pfeifer - 04:38 - 2 Dez. 2019