ES ph noise 1214031005

Dear Hanna,

basically the problem persists, the head of the lab told me she is performing bloodgas analysis and sample handling for the last 29 years without issure.
They are using 100 ul with 100 I.U./mL Heparin capillaries.
The patient's blood is drawn directly next to the instrument

Mainly they the problem is that the first try is a failure and consistently the second try is good

attached are some printouts

best regards
Tobias


by Tobias Pfeifer - 11:49 - 27 Nov 2019

Follow-Ups

  • RE: ES ph noise 1214031005

    Dear Tobias,

     

    Thank you for the update.

     

    How about changing the sampler assembly? Did they try it?

     

    Please let me know if changing the electrodes helped.

     

    Best regards,

     

    Hanna Parhizkar | Technical Support  | hparhizkar@medicacorp.com

    Medica Corporation | 5 Oak Park Drive, Bedford, MA 01730 | www.medicacorp.com

     

    +1 781 275 4892 Main

    +1 781 541 7403 Direct                    

    +1 781 275 2731 Fax

     

     

     

     

     

    From: FAMECO \ Tobias PFEIFER <tobias.pfeifer@fameco.eu>
    Sent: Monday, December 2, 2019 10:46 AM
    To: MEDICA TechnicalSupport <medicatechnicalsupport@fameco.odoo.com>
    Subject: Re: ES ph noise 1214031005

     

    Dear Hanna,

     

    The probe wiper is “new” changed on 10 NOV

     

    The problem was persistent with two ES instruments, I guess the customer wants to try and exchange the electrodes next 

    Tobias Pfeifer 

    --

    Image removed by sender.

    Tobias Pfeifer | Geschäftsführung


    T: +49 89 248805810 | F: +49 89 248805819

    E: tobias.pfeifer@fameco.eu | W: https://fameco.eu/

    FAMECO GmbH | Geschäftsführer: Tobias Pfeifer, Thomas Andersen

    Neumarkter Str. 34, 81673 München

    USt-IdNr. DE323800861 | Amtsgericht München HRB 248816

    The company accepts no liability for the content of this email, or for the consequences of any actions taken on the basis of the information provided, unless that information is subsequently confirmed in writing. If you are not the intended recipient you are notified that disclosing, copying, distributing or taking any action in reliance on the contents of this information is strictly prohibited.



    Am 02.12.2019 um 16:09 schrieb Hanna Parhizkar <HParhizkar@medicacorp.com>:

    

    Dear Tobias,

     

    Thank you for your email. I apologize for the delay in responding. Medica was closed for Thanksgiving holiday on Thursday and Friday.

     

    1. When was the last time that you changed the probe wiper? Could you please change the probe wiper?

    2. After changing the probe wiper if you still get the error, please change the sampler assembly .

     

    Please let us know how it goes.

     

    Hanna Parhizkar | Technical Support  | hparhizkar@medicacorp.com

    Medica Corporation | 5 Oak Park Drive, Bedford, MA 01730 | www.medicacorp.com

     

    +1 781 275 4892 Main

    +1 781 541 7403 Direct                    

    +1 781 275 2731 Fax

     

     

    Image removed by sender. cid:image002.png@01D5A8F7.DAAF07B0

     

     

     

    From: Tobias Pfeifer <tobias.pfeifer@fameco.eu>
    Sent: Wednesday, November 27, 2019 11:49 AM
    To: MEDICA TechnicalSupport <medicatechnicalsupport@fameco.odoo.com>
    Subject: ES ph noise 1214031005

     

    Dear Hanna,

    basically the problem persists, the head of the lab told me she is performing bloodgas analysis and sample handling for the last 29 years without issure.
    They are using 100 ul with 100 I.U./mL Heparin capillaries.
    The patient's blood is drawn directly next to the instrument

    Mainly they the problem is that the first try is a failure and consistently the second try is good

    attached are some printouts

    best regards
    Tobias


    --

    Image removed by sender. Image removed by sender.

    Tobias Pfeifer | Geschäftsführung

    T: +49 89 248805810  | F: +49 89 248805819

    E: tobias.pfeifer@fameco.eu | W: https://fameco.eu/

    FAMECO GmbH | Geschäftsführer: Tobias Pfeifer, Thomas Andersen

    Neumarkter Str. 34, 81673 München

    USt-IdNr. DE323800861 | Amtsgericht München HRB 248816

    The company accepts no liability for the content of this email, or for the consequences of any actions taken on the basis of the information provided, unless that information is subsequently confirmed in writing. If you are not the intended recipient you are notified that disclosing, copying, distributing or taking any action in reliance on the contents of this information is strictly prohibited.

     

    Sent by FAMECO GmbH using Odoo.

    --
    Hanna Parhizkar

    Sent by FAMECO GmbH using Odoo.

    <Changing the sampler.pdf>


    --

    Image removed by sender.

    Tobias Pfeifer | Geschäftsführung

    T: +49 89 248805810  | F: +49 89 248805819

    E: tobias.pfeifer@fameco.eu | W: https://fameco.eu/

    FAMECO GmbH | Geschäftsführer: Tobias Pfeifer, Thomas Andersen

    Neumarkter Str. 34, 81673 München

    USt-IdNr. DE323800861 | Amtsgericht München HRB 248816

    The company accepts no liability for the content of this email, or for the consequences of any actions taken on the basis of the information provided, unless that information is subsequently confirmed in writing. If you are not the intended recipient you are notified that disclosing, copying, distributing or taking any action in reliance on the contents of this information is strictly prohibited.

     

    Sent by FAMECO GmbH using Odoo.


    by Hanna Parhizkar - 02:54 - 2 Dec 2019
  • Re: ES ph noise 1214031005
    Dear Hanna,

    The probe wiper is “new” changed on 10 NOV

    The problem was persistent with two ES instruments, I guess the customer wants to try and exchange the electrodes next 

    Tobias Pfeifer 

    --

    Tobias Pfeifer Geschäftsführung


    T: +49 89 248805810 | F: +49 89 248805819

    E: tobias.pfeifer@fameco.eu | W: https://fameco.eu/

    FAMECO GmbH | Geschäftsführer: Tobias Pfeifer, Thomas Andersen

    Neumarkter Str. 34, 81673 München

    USt-IdNr. DE323800861 | Amtsgericht München HRB 248816

    The company accepts no liability for the content of this email, or for the consequences of any actions taken on the basis of the information provided, unless that information is subsequently confirmed in writing. If you are not the intended recipient you are notified that disclosing, copying, distributing or taking any action in reliance on the contents of this information is strictly prohibited.


    Am 02.12.2019 um 16:09 schrieb Hanna Parhizkar <HParhizkar@medicacorp.com>:

    

    Dear Tobias,

     

    Thank you for your email. I apologize for the delay in responding. Medica was closed for Thanksgiving holiday on Thursday and Friday.

     

    1. When was the last time that you changed the probe wiper? Could you please change the probe wiper?

    2. After changing the probe wiper if you still get the error, please change the sampler assembly .

     

    Please let us know how it goes.

     

    Hanna Parhizkar | Technical Support  | hparhizkar@medicacorp.com

    Medica Corporation | 5 Oak Park Drive, Bedford, MA 01730 | www.medicacorp.com

     

    +1 781 275 4892 Main

    +1 781 541 7403 Direct                    

    +1 781 275 2731 Fax

     

     

    cid:image002.png@01D5A8F7.DAAF07B0

     

     

     

    From: Tobias Pfeifer <tobias.pfeifer@fameco.eu>
    Sent: Wednesday, November 27, 2019 11:49 AM
    To: MEDICA TechnicalSupport <medicatechnicalsupport@fameco.odoo.com>
    Subject: ES ph noise 1214031005

     

    Dear Hanna,

    basically the problem persists, the head of the lab told me she is performing bloodgas analysis and sample handling for the last 29 years without issure.
    They are using 100 ul with 100 I.U./mL Heparin capillaries.
    The patient's blood is drawn directly next to the instrument

    Mainly they the problem is that the first try is a failure and consistently the second try is good

    attached are some printouts

    best regards
    Tobias


    --

    Image removed by sender.

    Tobias Pfeifer | Geschäftsführung

    T: +49 89 248805810  | F: +49 89 248805819

    E: tobias.pfeifer@fameco.eu | W: https://fameco.eu/

    FAMECO GmbH | Geschäftsführer: Tobias Pfeifer, Thomas Andersen

    Neumarkter Str. 34, 81673 München

    USt-IdNr. DE323800861 | Amtsgericht München HRB 248816

    The company accepts no liability for the content of this email, or for the consequences of any actions taken on the basis of the information provided, unless that information is subsequently confirmed in writing. If you are not the intended recipient you are notified that disclosing, copying, distributing or taking any action in reliance on the contents of this information is strictly prohibited.

     

    Sent by FAMECO GmbH using Odoo.

    --
    Hanna Parhizkar

    Sent by FAMECO GmbH using Odoo.

    <Changing the sampler.pdf>

    by Tobias Pfeifer - 10:45 - 2 Dec 2019
  • RE: ES ph noise 1214031005

    Dear Tobias,

     

    Thank you for your email. I apologize for the delay in responding. Medica was closed for Thanksgiving holiday on Thursday and Friday.

     

    1. When was the last time that you changed the probe wiper? Could you please change the probe wiper?

    2. After changing the probe wiper if you still get the error, please change the sampler assembly .

     

    Please let us know how it goes.

     

    Hanna Parhizkar | Technical Support  | hparhizkar@medicacorp.com

    Medica Corporation | 5 Oak Park Drive, Bedford, MA 01730 | www.medicacorp.com

     

    +1 781 275 4892 Main

    +1 781 541 7403 Direct                    

    +1 781 275 2731 Fax

     

     

    cid:image002.png@01D5A8F7.DAAF07B0

     

     

     

    From: Tobias Pfeifer <tobias.pfeifer@fameco.eu>
    Sent: Wednesday, November 27, 2019 11:49 AM
    To: MEDICA TechnicalSupport <medicatechnicalsupport@fameco.odoo.com>
    Subject: ES ph noise 1214031005

     

    Dear Hanna,

    basically the problem persists, the head of the lab told me she is performing bloodgas analysis and sample handling for the last 29 years without issure.
    They are using 100 ul with 100 I.U./mL Heparin capillaries.
    The patient's blood is drawn directly next to the instrument

    Mainly they the problem is that the first try is a failure and consistently the second try is good

    attached are some printouts

    best regards
    Tobias


    --

    Image removed by sender.

    Tobias Pfeifer | Geschäftsführung

    T: +49 89 248805810  | F: +49 89 248805819

    E: tobias.pfeifer@fameco.eu | W: https://fameco.eu/

    FAMECO GmbH | Geschäftsführer: Tobias Pfeifer, Thomas Andersen

    Neumarkter Str. 34, 81673 München

    USt-IdNr. DE323800861 | Amtsgericht München HRB 248816

    The company accepts no liability for the content of this email, or for the consequences of any actions taken on the basis of the information provided, unless that information is subsequently confirmed in writing. If you are not the intended recipient you are notified that disclosing, copying, distributing or taking any action in reliance on the contents of this information is strictly prohibited.

     

    Sent by FAMECO GmbH using Odoo.


    by Hanna Parhizkar - 10:08 - 2 Dec 2019